THS – FI022GB2K8v3 --- V - MAINTENANCE
82
Remote Possible settings Standard setting Parameter
-
CR
- -
Related to the current
product
Description Models
Reset of the current product counters
Resets the alarm and object counters for the current product.
All models
Remote Possible settings Standard setting Parameter
-
AR
- -
Global
Description Models
Reset total alarm counter
All models
Remote Possible settings Standard setting Parameter
-
OR
- -
Global
Description Models
Reset total pack transit counter
All models
Remote programming via serial link
Programming is carried out through a remote computer fitted with an RS232 or RS485 interface, using a standard
communications program (e.g. Windows™ Terminal) or the CEIA MDScope program, available on request.
Communication parameter settings:
• Baud rate = 9600 bit/s (to be set on the computer)
• 8 data bits
• NO parity
• 1 stop bit
• no communication protocol
• CR -> CR+LF
• Local echo = OFF
• Terminal: VT 100 (ANSI)
Entering remote programming
Connecting to a metal detector (via RS 232)
- Run the communications program
- Enter the password and key ENTER. The command
prompt will appear
(password) <ENTER>
#
Note: each password accesses programming at the appropriate level (operator, supervisor, …)
Long-distance connection to a single metal detector (via RS 485)
Identical to the above.
Network connection (via RS 485)
Using RS 485 transmission, several metal detectors can be connected in a network. In this case, the PC acts as master
and the metal detectors as slaves
. So as to allow each individual metal detector to be identified, the following steps must
be taken:
Set a password on each metal detector (N.B.: the password can be the same if so desired).
Set a different
Line ID identification code for each metal detector (line ID parameter ID=1001, 1002, 1003, .).