THS – FI022GB2K8v3 --- V - MAINTENANCE
81
Remote Possible settings Standard setting Parameter
SS diameter
SD
0.0 – 25.0 -
Global
Description Models
Minimum detectable diameter for stainless steel sample sphere
Datum that must be set for inclusion in the test report.
Note: The value is printed in the Q.C. report and does not effect the sensitivity.
All models,
except THS/MN
Remote Possible settings Standard setting Parameter
Non_FE diameter
ND
0.0 – 25.0 -
Global
Description Models
Minimum detectable diameter for diamagnetic metal sample sphere
Datum that must be set for inclusion in the test report.
Note: The value is printed in the Q.C. report and does not effect the sensitivity.
All models,
except THS/MN
Commands accessible only in Remote Programming
Remote Possible settings Standard setting Parameter
-
PE
- -
Global
Description Models
Exit from remote programming
In case the remote programming is not used for 3 minutes, the PE command is sent automatically by the THS
All models
Remote Possible settings Standard setting Parameter
-
PV
- -
Global
Description Models
Software version
Provides the software version loaded on the THS.
#PV <ENTER>
THS/STD V3.390a – ALM V1.060
All models
Remote Possible settings Standard setting Parameter
-
HE
- -
Global
Description Models
Help
Provides the list of commands available, their meanings and their current settings
All models
Remote Possible settings Standard setting Parameter
-
DP
- -
Global
Description Models
Cancellation of all products
The command delete ALL products defined by the user, leaving only the “Default” product.
All models
Remote Possible settings Standard setting Parameter
-
DA
DD/MM/YY -
Global
Description Models
Current date insertion
DD = Day, MM = Month
YY = last two figures of the year
#DA=18/02/04 <ENTER>
#
All models