THS – FI022GB2K8v3 --- V - MAINTENANCE
78
Print menu
Remote Possible settings Standard setting Parameter
Print parameters
PT
- -
Global
Description Models
Print list of parameters with their current settings
All models
The data are sent to the RS232 interface, if the
command is activated by local programming or
via RS232.
The data are sent to the RS485 if the command is
request via RS485.
Example:
#PT<ENTER>
PN P2
SE 203
TP 0
BA MEDIUM
AM 2
EM B
ED 500
RD OFF
BS 250
...
JR ON
MR ON
TR ON
RT 0:10
TM 2:58
DA 01/03/00
SN 999999
PV THS/3F V3.390a - ALM V1.060
FIN 013A---a - 001A----
#
Remote Possible settings Standard setting Parameter
Print buffer
PB
- -
Global
Description Models
Print list of memorised events
The data are sent to the RS232
interface, if the command is activated
by local programming or via RS232.
The data are sent to the RS485 if the
command is request via RS485.
The metal detector stores up to 1000
events. An event could be a certain
parameter modification (SE, TP ecc.),
a test result, an ejection (the result of
the ejection is memorized).
Example:
QUALITY REPORT 28/02/00 - 20:14:00
Serial number 999999
Line ID 1013
>Change product 28/02/00 - 20:13:13
P.Name P2
Operator ENGINEER
>Test Metal Detector 28/02/00 - 20:13:38
P.Name P2
Fe diameter 1.0
Test OK YES
Operator Q-C OPERATOR
>Test Metal Detector 28/02/00 - 20:13:49
P.Name P2
SS diameter 1.5
Test OK NO
Operator Q-C OPERATOR
>Statistics
END QUALITY REPORT
--------------------------------------------
All models