THS – FI022GB2K8v3 --- V - MAINTENANCE
71
Remote Possible settings Standard setting Parameter
Barcode enab.
BE
6 alphanum. characters -
Global
Description Models
Enabling code for operation with bar-code reader
6-character code which enables the THS to work with the bar-code reader: this is supplied by CEIA on delivery of an
upgrade kit.
All models,
except THS/G
Remote Possible settings Standard setting Parameter
Phcell position
PH
IN / OUT -
Global
Description Models
Position of the photocell relative to the probe
IN: photocell located at the entrance to the probe; OUT: photocell located at exit of the probe.
All models
Remote Possible settings Standard setting Parameter
Phcell-MD dist.
PD
0 – 2000 mm -
Global
Description Models
Distance between metal detector and photocell
The distance must be measured from the center of the photocell to the side of the probe.
All models
Remote Possible settings Standard setting Parameter
K trans.
KT
0.000 – 9.999 / F -
Global
Description Models
Transmission constant
KT= (transit speed)/(motor frequency) ; KT=0.000: no inverter
Note: when the value of KT is altered, parameters BL and BM are automatically set to nominal
values 20 x KT and MI x KT respectively. BL and BM can nevertheless be altered manually as
well.
Parameter KT must be set to “F” if the system configuration includes an MDT card.
All models,
except THS/G
Remote Possible settings Standard setting Parameter
K encoder
KE
0 – 1000 pulses / revol. 0
Global
Description Models
Encoder constant
Number of pulses per revolution. If set to 0 the encoder is not present.
All models,
except THS/G
Remote Possible settings Standard setting Parameter
Diameter
DI
10 – 250 mm -
Global
Description Models
Diameter of encoder roller
Diameter of encoder roller. Correlated parameter: K encoder.
All models,
except THS/G
Remote Possible settings Standard setting Parameter
TP adjustment
TA
0 – 1799 -
Global
Description Models
Correction of the received signal analysis parameter
This parameter is specific to each device. This parameter must be automatically set up using the TL parameter in
case the SCD card is replaced.
All models,
except THS/MN
Remote Possible settings Standard setting Parameter
TA autolearn
TL
ON / OFF -
Global
Description Models
Autolearn of parameter TA
Select ON to activate the autolearn procedure (see TA parameter) and exit from programming.
Note: TL=OFF automatically at the end of procedure. See paragraph “Replacement of SCD card
”
All models,
except THS/MN