Manuale operativo
CHORUS
110 Rev. 0 del 06/2012
The errors from 138 to 143 may be found during the optic check, and indicate poor function of the
offset regulation of the channel indicated
144 Channel #1 dark
145 Channel #2 dark
146 Channel #3 dark
147 Channel #4 dark
148 Channel #5 dark
149 Channel #6 dark
The errors from 144 to 149 may be found during the optic check, and indicate poor function of the Dark
reading of the channel indicated
150 Channel #1 air (650)
151 Channel #2 air (650)
152 Channel #3 air (650)
153 Channel #4 air (650)
154 Channel #5 air (650)
155 Channel #6 air(650)
156 Channel #1 air (610)
157 Chanell #2 air (610)
The errors from 150 to 155 may be found during the optic check, and indicate poor function of the Air
reading of the channel indicated.
5.4.2.4 Identification errors
Represent the errors detected during the identification procedure (the corresponding strips will not be
processed
STR EXP
Description: the validity of the strip is expired.
STR CO
Description: the strip is to be incompatible with the others present in the tray
STR RD
Description: the strip barcode is illegible.
STR ID
Description: the test code of the strip is not recognized
NO CALIB
Description: the corresponding strips is part of a non calibrated batch
CAL POS
Description: the calibrator is positioned after a strip with the same code. The calibrator must
precede the relative strip on the tray