Manuale operativo
CHORUS
112 Rev. 0 del 06/2012
DISPSYR#1
Description: the 250ml syringe did not dispense the correct quantity of liquid because of a
loss of steps of the instrument
The error code is S1
DISPSYR#2
Description: the 25ml syringe did not dispense the correct quantity of liquid because of a loss
of steps of the instrument.
The error code is S2
ANNULL
Description: warning of not execution of the tests due to an error of the instrument
5.4.2.6 Analytical errors
Represent the errors in the calculation of the test results
SeeManual0
Description: Error in the optic reading. Optic reader not calibrated or broke: execute optic
calibration
SeeManual1
Description: the kinetics of the processed sample does not correspond to the kinetics
expected of the test, because the first reading is out of range: dilute the sample
and repeat the test and/or contact customer support.
SeeManual2
Description: the kinetics of the processed sample does not correspond to the kinetics
expected of the test because the trend is descending: dilute the sample and
repeat the test and/or contact customer support.
SeeManual3
Description: the kinetics of the process sample does not correspond to the kinetics expected
of the test because the trend is increasing excessively : dilute the sample and
repeat the test and/or contact customer support.
SeeManual4
Description: The initial CFT reading is too high: change the strip because of hemolyzed or not
properly mixed red blood cells and/or contact customer support.