Manuale operativo
CHORUS
116 Rev. 0 del 06/2012
If the Skip key is pressed, the instrument will not
permit you to start the priming of the autoimmunity
buffer solution which was selected as current
buffer.
It will not be possible to start a working cycle until
the buffer solution is identified correctly.
fig. 121
5.5.3 I
DENTIFICATION OF THE
B
UFFER
S
OLUTION FOR
A
UTOIMMUNITY
A window is displayed (
f
ig
.
1
22
) which requests
removal of the probe of the buffer solution for
autoimmunity (this window will not appear when the
procedure is started automatically after the
extraction of the probe from the solution of the
tank).
Using the Skip key, it is possible to skip this check
(fig. 125).
fig. 122
As soon as removal of the probe is detected, the
following window (fig. 123) will request the reading
of the bar code. Whenever a bar code reader is not
available, it is possible to enter the code manually
by pressing the Edit key (code = 42462002).
If the text Wrong auto. buff. bar codeis displayed
after the introduction of the code, this indicates that
you are trying to enter a different bar code than
expected by the instrument.
Using the Skip key, it is possible to skip this check
(fig. 125).
fig. 123
Subsequently, you will be requested to insert the
probe into the tank (not empty) of the buffer
solution that was previously identified (fig. 124).
Using the Skip key, it is possible to skip this check
(fig. 125).
fig. 124