Manuale operativo
CHORUS
Rev. 0 del 06/2012 111
CAL NR
Description: more than one calibrator was found for the same batch
CAL/STR
Description: the calibrator is associated to a different test or batch
QC/STR
Description: the control serum is associated to a different test or batch
STR REP
Description: the strip was already processed during the previous run
STR BUF
Description: the strip is incompatible with the buffer solution present in the hydraulic circuit
5.4.2.5 Errors during the cycle
Represent the errors detected during the cycle (the corresponding strips will be aborted). The error
regarding the aborted strip will be reported (by means of a code) in the OP column, in the printed
report and saved in the archives.
CUV MIS
Description: during the pre-heating procedure well 5 or 6 or 7 is not detected in the strip. The
code of the error is CM.
SMP QT
Description: the amount of serum dispensed in the reaction well is insufficient
The error code is SQ.
DRILL#1
Description: dispensator 1 has not perforated the strip well correctly
The error code is D1.
DRILL#2
Description: dispensator 2 has not perforated the strip well correctly
The error code is D2.
DISP#1
Description: dispensator 1 has not transferred the correct amount into the measuring well
The error code is T1.
DISP#2
Description: dispensator 2 has not transferred the correct amount into the measuring well
The error code is T2.
LOW CAL
Description: the sample is a calibrator and its value is below the minimum acceptable limit
HIGH CAL
Description: the sample is a calibrator and its value is above the maximum acceptable limit
CAL OVR
Description: the sample is inserted in a run with its relative calibrator on which the test has
failed.