Manuale operativo
CHORUS
Rev. 0 del 06/2012 79
The pediatric codes are shown in blue. (This characteristic is not used at the moment, see 4.4.1)
4.8.3 P
RINTED REPORT
Below is an example of the printout of the analysis results:
CHORUS DIESSE Diagnostica Senese SpA
2006/11/01 12:00 Rel. 2.09r21 Meth. 11 S/N: 1000/10/2006
Nr Smp Code Test R Value MU Min Max Lot
---------------------------------------------------------------------
01 Qc TX-av TX-av H 50.00 % 40 60 401
02 Qc TX-av TX-av L !25.00 % 40 60 401
03 Cal RV-G RV-G - 200 412
T h e f o l l o w i n g s a m p l e h a s b e e n p r o c e s s e d w i t h a n e x p i r e d d e v i c e :
04 Sample #1 TX-av L 20.00 % 401
05 T Sample #2 ?TX-av N.C. % 401
06p Sample #3 TX-av ManualOp0 % 401
07 Sample #4 TX-av N.C % 401
08 Qc RV-G RV-G N !7.5 IU/ml 10.0 50.0 412
09 Sample #5 ?RV-G D 10.0 IU/ml 412
10 Sample #6 TX-G T_ERR 410
fig. 90
The meaning of columns is the following:
Nr:
Position of the strip on the tray. If followed by the letter “p”, it is a pediatric sample (this characteristic is not used
at the moment, see 4.4.1)
Smp code:
Sample code. If preceded by the letter T, the strip has been processed with a temperature error and the result
could therefore not be correct
Possible expired strips will be preceded by the phrase “ The following sample is process with a expired strip”
Test:
Test performed. If preceded by the character ? , the strip code was not identified by the barcode reader and was
inserted manually (therefore the system cannot guarantee its validity)
R (reference):
This represents the interpretation of the numeric result compared to the cut-off.
Possible values are: P (positive), N(negative), D (doubtful).