Manuale operativo
CHORUS
Rev. 0 del 06/2012 87
4.11.1 I
NFORMATION ON THE STATE OF THE INSTRUMENT
Strip temp: Temperature measured inside the thermostatic chamber
Instrument temp.: Temperature measured inside the instrument
Wash. buf. level (infectivity or autoimmunity)/ Clean. sol. level: Level of liquid in the buffer/cleaning solution
tanks.
The values may be:
. 75%-100% the level is between 75% and 100%
50%-75% the level is between 50% and 75%
25%-50% the level is between 25% and 50%
0%-25% the level is between 0% and 25% (it is not possible to start the run)
non id. the tank was not correctly identified (in red)
Empty: the tank is completely empty (it is not possible to start the run)
Optic calibration:
Indicates (in red) when optic calibration is necessary (if more than 24 hours have passed or if the previous
calibration was interrupted)
Priming:
Indicates (in red) when priming is necessary (if the circuit contains liquid other than the Buffer or if the previous
priming procedure was interrupted, or in case more than 6 hours have passed since the last priming)
Hydraulic Check:
Indicates (in red) when it is necessary to perform a hydraulic check (if more than 24 hours have passed or if the
previous check was interrupted)
Liquid in the circuit:
Indicates which liquid is present in the Buffer circuit.
The possible values are:
Air indicated in red
Infect. Washing buffer Washing Buffer for infectivology tests (in red if different from the Washing
buffer selected)
Autoimm. Washing buffer Washing Buffer for autoimmunity tests (in red if different from the
Washing buffer selected)
Washing sol. indicated in red
Sanit. sol. indicated in red
ATTENTION
When these indications are signaled in red, priming is necessary.