Manuale operativo
CHORUS
32 Rev. 0 del 06/2012
No further controls will be performed on the lid.
An error message (warning) will appear in the first error window, at the end of the check.
If the window should close due to time-up, the system check will proceed.
When necessary, opening/closing of the lid will be requested again, to control the error.
The error (warning) remains until the lid is correctly positioned in a subsequent control.
Note:
If the Skip key is pressed in any part of the program except in the initial check, the error remains until a
new initial check is executed.
ATTENTION
The lid is checked again when the R-List is entered, and if this step is skipped (using the Skip key or
due to time-out), the warning signal will remain on the status bar throughout the cycle, indicating the
risk for the operator and for performing the test itself.
ATTENTION
It is important to stress that if the operator continues to run the instrument with the lid warning signal
on, he does so entirely at his own risk.
3.4.4 C
ONTROL OF THE MOTORS
The control procedure for the motors moves each one into its respective working position.
In case a motor can not be positioned correctly, a relative error will appear in the errors window and
the instrument check will be interrupted.
3.4.5 C
ONTROL OF THE STATE OF THE INSTRUMENT
The instrument check continues with the control of the following pendings:
• Pending of the hydraulic system
that is activated whenever
- the last hydraulic check was unsatisfactory,
- more than 24 hours have passed since the last check,
and enables execution of the initial hydraulic check.
• Pending of the priming of the washing buffer circuit
that is activated whenever
- the washing buffer is not present in the circuit,
- more than 6 hours have passed since the last priming check or last cycle,
- previously the hydraulic check was executed,
and thus enables the execution of the priming of the buffer circuit.
• Pending of the optic system
is activated whenever
- the last optic calibration check was not concluded satisfactorily,
- more than 24 hours have passed since the last optic calibration.