Manuale operativo
CHORUS
20 Rev. 0 del 06/2012
3 PREPARATION OF THE INSTRUMENT
This chapter includes unpacking instructions and is inserted on the external cover of the transport box.
It has to be carefully read and the operations described are to be executed by specialized personnel of
DIESSE Diagnostica Senese SpA or personnel authorized by DIESSE.
3.1 HANDLING
3.1.1 T
RANSPORT AND
S
TORAGE
The packaging is a wooden box with the dimensions of 100 x 100 x 100 cm on a pallet
During storage and transportation the indications of TOP and BOTTOM, printed on the outside, have to
be observed. The instrument can be transported in temperature conditions included between 0°C and
50°C. Do not leave the boxes in rainy or humid conditions. In case the instrument has been exposed to
a temperature below 10°C for 24 hours or longer, the installation personnel should check to ensure
there is no condensation in the encoding systems, in the optical devices and in the moving parts of the
instrument, and should ensure that the instrument will not be turned on again before it is left at room
temperature for an hour.
3.1.2 U
NPACKING
To open the box and the other packaging of the instrument, the following utensils are necessary:
- Two allen keys of 2,5 and 3,0 mm (not provided).
- A big cross screwdriver (not provided).
After ensuring that the packing is not ruined and that the TOP/BOTTOM positions are correct, proceed
as follows:
fig. 7
Unscrew the 4 screws on the cover
Remove the cover