Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 129
Sensor Data - Min.
Temp.Carrier
read only The lowest carrier tube temperature measured appears on the display.
! Note!
This function is not available for the Promass E measuring device.
Sensor Data -
Max. Temp.Carrier
read only The highest carrier tube temperature measured appears on the display.
! Note!
This function is not available for the Promass E measuring device.
Sensor Data -
Temp.Unit
read only Displays the unit of the temperature.
! Note!
The unit is taken from the parameter "System Unit - Temperature"
( → Page 117).
Simulation -
Measurand
AUTO - OOS Use this parameter to set all inputs, outputs and totalizers to their defined
flow-response modes, in order to check whether they respond correctly.
During this time, the words "SIMULATION MEASURAND" appear on the
display.
Options:
OFF
MASSFLOW
VOLUMEFLOW
COR. VOLUMEFLOW
DENSITY
REFERENCE DENSITY
TEMPERATURE
Factory setting:
OFF
! Note!
The active simulation mode is passed on to subsequent function blocks via
the status condition (UNCERTAIN) of the process variable.
" Caution!
• The measuring device cannot be used for measuring while this simulation
is in progress.
• The setting is not saved if the power supply fails.
Simulation -
Value
Measurand
AUTO - OOS
! Note!
The entry is not active unless a measured variable was selected in the
"Simulation - Measurand" parameter.
Use this parameter to define a selectable value (e.g. 12 m
3
/s). This is used to
test the associated parameters in the flowmeter itself and downstream
function blocks.
User input:
5-digit floating-point number
Factory setting:
0 [unit]
" Caution!
The setting is not saved if the power supply fails.
"Flow" Transducer Block / base index 1400:
Parameter
Write access
with operating
mode
(MODE_BLK)
Description