DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-66 Section 3-8 - Mounting the Vivid i n/ Vivid q N on the SafeLock Cart (optional)
3-8-1 Mounting the Vivid i n and Vivid q N on the SafeLock Cart (cont’d)
2.) Push the scanner down and backwards, until it locks into place against the locking tooth.
An audible click will be heard as the scanner locks into position. There will now be a gap the width
of one finger between the SafeLock Cart arm rest and the front of the scanner.
3.) If necessary, perform the alignment procedure as described in 3-8-2 "Adjusting the SafeLock Cart
Locking Mechanism" on page 3-68.
4.) Connect the LAN Cable from the cart to the LAN connector on the Vivid i n and Vivid q N rear panel.
5.) Connect the Docking Connector to the rear of the scanner, as shown below.
Figure 3-63 Vivid i n and Vivid q N Scanner Mounted on SafeLock Cart
Figure 3-64 Vivid i n and Vivid q N Rear Panel
Figure 3-65 Connecting the Docking Connector to the Rear of Scanner
Docking Connector
LAN Connector
USB Port
Docking Connector