GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 10 - Care and Maintenance 10-9
Section 10-5
System Maintenance
10-5-1 Preliminary Checks
The preliminary checks take approximately 15 minutes to perform. Refer to the Vivid i n/Vivid q N User
Manual whenever necessary.
Table 10-4 System Preliminary Checks
St
ep Item Description
1
Ask & Listen Ask the customer if they have any problems or questions about the equipment.
2
Paperwork
Fill in the appropriate details at the top of the Vivid i n/ Vivid q N System Inspection Certificate
(see page 10-34).
Note all probes and system options.
3
Power-up
Turn the system power ON and verify that all fans and peripherals turn on. Watch the displays
during power up to verify that no warning or error messages are displayed.
4
Probes Verify that the system properly recognizes all probes.
5
Displays Verify proper display on the monitor and touch panel.
6
Presets Back-up all customer presets on a CD-RW.