DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-58 Section 8-3 - Internal Component Replacement Procedures
8-3-9 LCD Flex Cable Replacement Procedure
8-3-9-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Flex Cable replacement
procedure.
8-3-9-2 Time Required
15 minutes
8-3-9-3 Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-9-4 LCD Flex Cable Removal Procedure
1) Disconnect the AC cable and battery.
2) Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal
Procedure on page 8 - 9.
3) Remove the LCD Display Frame as described in the LCD Display Frame Removal Procedure on
page 8 - 27.
4) Remove the LCD Display as described in the LCD Display Removal Procedure on page 8 - 53.
5) Using both hands, carefully lift the LCD Display upwards and remove it from the Rear Cover & Latch
Assembly, as shown in Figure 8-68.
NOTE: Take care not to damage the LCD Flex ribbon cable when performing this step.
Flex Cable Replacement Procedure
Figure 8-68 Removing the LCD Display