DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-3
7-2-3 Accessing the Diagnostic Test Window
The diagnostic tools are accessed from 2D-Mode by simultaneously pressing <ALT+Config> on the
alphanumeric keyboard. The Diagnostic Test window is displayed, as shown below:
Figure 7-1 Diagnostic Test Window
NOTICE
IMPORTANT - The user-friendly Diagnostic Test Window displays dynamically-updated information.
The Data Flow Map area on the right is designed to provide an instant visual representation of the system
components; the color illumination of graphic buttons (light blue; light gray) and status indicators (green;
red; yellow) indicates which component, or set of components, is currently undergoing diagnostic testing
(see the examples in Figure 7-2 on page 7-4).
In this way, it is easy to visually follow the test sequence and progress of a specific test.
For a detailed explanation of the Diagnostic Test Window, refer to Table 7-2 on page 7-5.
Data Flow
Diagnostic
Status Area
Current Test
Test
Tree
Controls &
Map
Instructions
Special