8-130 Section 8-5 - Software Loading
8-5-7 Installing Software only from DVD
NOTE: During software installation, all peripherals should be removed - including PCMCIA devices (but
excluding the DVD\CD-RW driver).
1) Connect the DVD\CD-RW to the USB port at the rear of the system.
2) Connect power to the DVD\CD-RW drive.
3) Place the DVD containing the full software version into the DVD\CD-RW drive and close the tray.
4) Turn ON the Vivid i n/ Vivid q N system and allow it to boot-up from the DVD.
The system powers ON and when system boot-up from the DVD is complete, the following options
are displayed in the DOS screen, prompting you to select the required option:
5) Press 2 to perform installation of Vivid i n/ Vivid q N Application Software only from DVD.
The system will continue to boot up and the following screen is displayed.
NOTICE
IMPORTANT It is recommended to store partition C:\ by using the Rollback tool (described in the
Software Roll-back Procedure on page 8 - 139) after full software installation or upgrade.