DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-66 Section 8-4 - SafeLock Cart Components Replacement
8-4-5 AC Power Cable Bumper Protector Replacement Procedure
8-4-5-1 Tools
Appropriate Phillips screwdriver and Loctite glue.
8-4-5-2 AC Power Cable Bumper Protector Removal Procedure
1) From the rear of the SafeLock Cart, and using two hands, lift the AC Power Cable Bumper Protector
upwards to release it from the shoulder screws - (refer to Figure 8-77 and Figure 8-75) .
2) Remove the Bumper Protector from the Cart.
8-4-5-3 AC Power Cable Bumper Protector Installation Procedure
NOTE: If this is a first-time installation of the bumper protector, it is necessary to remove the two existing Phillips
screws located on the rear of the SafeLock Cart (AC Cable connection area), as shown in Figure 8-75.
1.) Apply a little Loctite glue on the threads of each of the two shoulder screws (Figure 8-76).
Figure 8-75 Screws for Removal on First-time Installation
Figure 8-76 Shoulder Screws
Apply Loctite glue here