GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-45
8-3-4 BIOS Battery Replacement
NOTE: Since the BIOS battery is not available as a spare part (due to shelf life), it should be obtained locally.
The battery model is CR2032.
8-3-4-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the BEP replacement procedures.
8-3-4-2 Time Required
15 minutes
8-3-4-3 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-4-4 BIOS Battery Removal Procedure
1) Remove the BEP, as described in the BEP Removal Procedure on page 8 - 38.
2) Take out the BIOS battery located on the bottom of the BEP, as shown in Figure 8-54.
8-3-4-5 BIOS Battery Installation Procedure
NOTE: Before proceeding, refer to the Preparation Procedure on page 8 - 43.
1) Fit a new BIOS battery in position on the BEP - refer to Figure 8-54.
2) Re-install the BEP, as described in the BEP Installation Procedure on page 8 - 42.
3.) Turn ON the system and boot-up.
4.) Adjust to the correct date and time settings - see Date and Time Adjustments on page 3 - 71.
5.) Proceed to perform the following functionality test:
- Battery Parameters Test on page 7 - 102
Figure 8-54 BEP - Location of BIOS Battery
BIOS Battery