DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-72 Section 8-4 - SafeLock Cart Components Replacement
8-4-8 AC Distribution Assembly Replacement Procedure
8-4-8-1 Tools
Appropriate flat and Phillips screwdrivers and a wire cutter
8-4-8-2 Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-8-3 AC Distribution Assembly
Removal Procedure
1) Turn OFF power to the SafeLock Cart and disconnect the AC power cable from the AC input
assembly at the base of the cart (rear).
2) Remove the Upper Cover (Rear), as described in the Upper Cover (Rear) Removal Procedure on
page 8 - 64.
3) Remove the Lower Cover (Rear), as described in the Lower Cover (Rear) Removal Procedure on
page 8 - 71.
4) Remove the three retaining screws of the AC Assembly Cover and remove the cover - Figure 8-83
below.
The AC Distribution Assembly is now exposed, as shown in Figure 8-84.
5) Cut all cable ties holding the cables in position.
6) Unscrew the metal cable fastener from the P1 cable connector (on the right) and remove the cable
from its socket - see Figure 8-84.
7) Disconnect the peripheral power cable from the AC Distribution Assembly - see Figure 8-84.
Figure 8-83 AC Distribution Assembly and Mounting Bracket
Retaining Screw