7-20 Section 7-3 - Performing Full System Diagnostics
7-3-1 Accessing the Full System Diagnostic Options
In 2D-Mode, press <ALT +Config> on the alphanumeric keyboard. The Diagnostic Test window is
displayed, as shown in (Figure 7-1).
7-3-1-1 Full System Test (FE and BEP)
NOTE: For a description of this test, refer to the Full System Test (FE and BEP) on page 7 - 10.
NOTE: The Full System Test (FE and BEP) performs a full sequence of complete system tests (automatically)
and requires approximately 10 minutes for completion.
1.) In the Diagnostic Test window, trackball to Full System Test (FE and BEP) in the list of tests
available for selection from the Diagnostic Test Tree (see Figure 7-5 below) and press Set. The Full
System Test (FE and BEP) name is displayed under Execute, and all system components are
highlighted blue in the Schematic View, as shown below:
Figure 7-5 Full System (FE and BEP) Test
2.) As indicated in the Special Instructions area, make sure the system is calibrated and disconnect
any probe currently connected to the system.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may