EasyManua.ls Logo

GE Vivid i n - Front Wheel Replacement Procedure

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-93
8-4-19 Front Wheel Replacement Procedure
NOTE: This section describes the replacement procedure for a SafeLock Cart front wheel. The brake on each
front wheel is an integral part of the wheel housing.
8-4-19-1 Tools
Appropriate flat screw driver and open wrench.
8-4-19-2 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-19-3 Front Wheel Removal Procedure
1) Disconnect mains power from the AC adapter.
2) Disconnect the AC adapter and all cables and peripherals from the Vivid i n/Vivid q N portable
ultrasound scanner.
3) Remove the Vivid i n/ Vivid q N and all peripherals from the SafeLock Cart.
4) Carefully lay the SafeLock Cart on its back.
5) Loosen and remove the four bolts that secure the front wheel to the SafeLock Cart chassis
(see Figure 8-106).
6) Remove the wheel from the SafeLock Cart.
Figure 8-106 Front Wheel Bolted to SafeLock Cart Chassis
Two front wheel bolts
shown
Break