DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-134 Section 8-5 - Software Loading
- Calibration Tests on page 7 - 25
- Full System Test on page 7 - 10
8-5-8 Installing Software only or Patch
1) Connect the DVD\CD-RW to the system’s USB port.
2) Connect the DVD\CD-RW power supply to the mains power.
3) Place the CD with Patch or Software Only CD into the tray and close.
4) Turn ON the system.
During boot-up, system will prompt to install software or continue booting into regular mode.
5) Click on Install Software and follow the on-screen instructions.
6.) When done, the Installation Complete message is displayed (refer to Figure 8-129 on page 8-133).
7) Move the cursor over the displayed message window and press the SET button.
8.) Press any key to re-boot the system.
The system re-boots into Normal Scanning mode.
9.) Press Config and then About.
The newly-installed software version is now listed under the System Version tab (refer to
Figure 8-130 on page 8-133).
10.)Remove the CD from the DVD\CD-RW drive and close the tray.
11.)Proceed to perform the following functionality tests:
- Software Configuration Checks on page 4 - 37
- Calibration Tests on page 7 - 25
- Full System Test on page 7 - 10
NOTICE
IMPORTANT It is recommended to store partition C:\ by using the Rollback tool (described in the
Software Roll-back Procedure on page 8 - 139) after full software installation or upgrade.