GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-110 Section 3-10 - Connectivity Setup
3-10-4 Set Up Connection to a DICOM Server in a Network
In this case, the Vivid i n/ Vivid q N is configured to work with a DICOM server in a network environment.
Images are first saved on the local image buffer on the scanner. At the end of the examination, the
images are sent to the DICOM server via a DICOM spooler.
This scenario requires that the scanner is configured to be connected to the DICOM server as described
below.
3-10-4-1 Overview
To work against the DICOM server, the following information has to be entered in the scanner:
• DICOM server IP address, subnet mask and eventually the gateway.
• DICOM server port number.
• DICOM server AE title (the server application’s name).
• Name of Device.
3-10-4-2 DICOM Server IP Address Setting on the Scanner
1.) Press CONFIG
(F2) and log on as Adm, as described in Log On to the System as ADM on page 4-10.
2.) Select CONNECTIVITY
(in the lower part of the window).
3.) Select the DATAFLOW
tab.
4.) Select the arrow to the right of the Name field to list all dataflows in a pull-down menu, see
Figure 3-115.
5.) From the Dataflow Name pull-down list, select the dataflow you want to configure - see Figure
3-115.
Figure 3-115 Select Dataflow
Arrow
Select dataflow from
Dataflow Name pull-down
menu.