DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-92 Section 8-4 - SafeLock Cart Components Replacement
8-4-18 Probe Cable Hook Replacement Procedure
8-4-18-1 Tools
Appropriate Phillips screwdriver.
8-4-18-2 Probe Cable Hook Removal Procedure
1) From below the Probe Shelf, remove the Probe Cable Hook retaining screw (see Figure 8-105 on
page 8-92).
2) Gently pull the Probe Cable Hook upwards to remove it.
8-4-18-3 Probe Cable Hook Installation Procedure
1) Insert the Probe Cable Hook in its position on the Probe Shelf.
2) From below, return and replace the Probe Cable Hook retaining screw.
Figure 8-105 Probe Cable Hooks
Screw accessed from below