DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-70 Section 3-9 - Configuration
3-9-2 Vivid i n/ Vivid q N Configuration
3-9-2-1 Preparations
Press CONFIG
(F2) and log on as Adm (see Log On to the System as ADM on page 4-10).
3-9-2-2 Enter Location
Figure 3-68 Hospital and Department Name
Table 3-7 Location Name
Step Task Expected Result(s)
1.
Open the Configuration Window, see
Preparations.
If needed, select SYSTEM.
The System Settings Window is displayed.
2.
Select the Hospital field, see (a) in Figure 3-68, and type the
name of the hospital (max 64 characters).
The 24 first characters of this name are displayed on the scanning
screen’s title bar (after restart)
All 64 characters are displayed on the image properties on saved images
(after restart).
3.
Select the Department field, see (b) in Figure 3-68, and type the
name of the department (max 64 characters).
This name will be displayed on the image properties on saved images as
soon as the unit has been restarted.
a
b