DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-87
8-4-15 Top Shelf Replacement Procedure
8-4-15-1 Tools
Appropriate Phillips screwdrivers, Allen key and open wrench.
8-4-15-2 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-15-3 Top Shelf Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid i n/Vivid q N portable ultrasound scanner.
2) Disconnect and remove all peripherals from the SafeLock Cart.
3) Remove the Upper Cover (Rear) as described in the Upper Cover (Rear) Removal Procedure on
page 8 - 64.
4) Remove the Lower Cover (Rear) as described in the Upper Cover (Rear) Removal Procedure on
page 8 - 64.
5) Remove the USB board as described in the USB Board Removal Procedure on page 8 - 80.
6) Remove the gas spring as described in Gas Spring Replacement Procedure on page 8 - 82.
7) Disconnect the two cables (P1 cable and peripheral power cable) from the top of the AC Input
Assembly as described in AC Distribution Assembly Removal Procedure on page 8 - 72 (refer to
step 6 and step 7).
8) Remove Main Harness, Peripheral power cable and disconnect the LAN cable.
9) Remove the six lower screws that retain the top Shelf - see Figure 8-100.
10) Remove the two rear screws that retain the top Shelf - see Figure 8-101.
11) Carefully remove the top shelf assembly.
Figure 8-100 Top Shelf - Lower Retaining Screws
Lower screws retaining the
Top Shelf - 6 shown