DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
10-12 Section 10-5 - System Maintenance
10-5-4 Cleaning
10-5-4-1 General Cleaning
Frequent and diligent cleaning of the Vivid i n/ Vivid q N ultrasound unit reduces the risk of spreading
infection from person to person, and also helps to maintain a clean working environment.
CAUTION: When performing Cleaning Procedures, to prevent the risk of system damage, always
observe the following precautions:
•Use only cleaning materials and solutions (such as Isopropyl Alcohol [IPA]) as
recommended in the procedures described in Table 10-8 on page 10-13.
Do not use any solutions or products not listed in the Vivid i n/Vivid q N User Manual.
Do not spray any liquid directly onto the Vivid i n/ Vivid q N covers, LCD Display or keyboard!
Do not allow any liquid to drip or seep into the system.
Prior to cleaning, turn OFF power to the system.