GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-64 Section 8-4 - SafeLock Cart Components Replacement
8-4-3 Upper Cover (Rear) Replacement Procedure
8-4-3-1 Tools
Appropriate Phillips screwdriver.
8-4-3-2 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-3-3 Upper Cover (Rear) Removal Procedure
Remove the retaining screws of the upper rear cover and while supporting the cover with one hand,
carefully remove the cover (see Figure 8-73).
8-4-3-4 Upper Cover (Rear) Installation Procedure
1.) Return the cover to the correct position on the rear of the SafeLock Cart (see Figure 8-73).
2) While holding the cover to prevent it from falling, return and fasten the two upper screws
(see Figure 8-73).
3) Fasten the four remaining screws that secure the upper cover (rear).
4) Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Figure 8-73 Upper Cover (Rear) Removal
Retaining Screw