DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-5
8-2-2 Battery Replacement Procedure
8-2-2-1 Tools
None required.
8-2-2-2 Time Required
1 minute
8-2-2-3 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-2-2-4 Battery Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid i n/Vivid q N portable ultrasound scanner.
2) Close the Vivid i n/ Vivid q N LCD cover, making sure both front latches are securely fastened.
3) Carefully turn the Vivid i n/ Vivid q N over, placing it face down.
4.) While holding the Vivid i n/ Vivid q N steady, pull the locking latch away from the battery and slide
the battery sideways out of the battery compartment to remove it completely from the system.
NOTE:
WHEN REMOVING A DEFECTIVE BATTERY, ENSURE THAT IT IS DISPOSED OF IN
ACCORDANCE WITH LOCAL REGULATIONS. ALTERNATIVELY, FORWARD IT TO
Figure 8-4 Closing Vivid i n/ Vivid q N LCD Cover
Figure 8-5 Sliding the Battery out of the Battery Compartment
LatchLatch