DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-2 Section 3-2 - Installation Reminders
Section 3-2
Installation Reminders
3-2-1 Average Installation Time
The Vivid i n/ Vivid q N installation and functional checkout will take approximately one hour; Vivid i n
or Vivid q Nconsoles with optional equipment may take slightly longer.
Once the site has been prepared, the average installation time required is shown in Table 3-2 below.
3-2-2 Installation Warnings
1.) The Vivid i n/Vivid q N portable ultrasound scanner weighs only 5 Kgs (11 lbs). (This is the weight
of the scanner without any optional accessories). However since the SafeLock Cart weighs
approximately 40 kg (88 lbs), two persons are always required to unpack it.
2.) There are no operator-serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
3-2-2-1 System Acclimation Time
Following transport, the Vivid i n/ Vivid q N system may be very cold, or hot. Allow time for the system
to acclimate before being switched ON. Acclimation requires 1 hour for each 2.5
o
C increment, when the
temperature of the system is below 10
o
C or above 35
o
C.
Table 3-2 Average Installation Time
Description
Average
Installation Time
Comments
Unpacking the scanner 20 minutes
Installing the scanner 30 minutes Time may vary, according to the required configuration
DICOM Option
(connectivity)
30 minutes Time may vary, according to the required configuration
InSite Option 30 minutes
CAUTION
Turning the system ON after arrival at the site - without allowing time for acclimation - may cause
system damage!
Table 3-3 Vivid i n and Vivid q N System Acclimation Time
ALLOWED TRANSPORTATION AND STORAGE TEMPERATURES OUT OF SPEC!
°C
60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40
°F
140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 -4 -13 -22 -31 -40
Hrs
864200000002468101214161820