GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
4-28 Section 4-3 - Functional Checks
4-3-1-11 Visual Test
4-3-2 Peripherals
4-3-2-1 VCR Test
Table 4-15 Visual Test - Recommended
Step Task Expected Result(s)
1
Insert 3S-RS probe.
2
Increase the gain to see the homogenous noise.
The probe ID is 3S-RS. No strikes or artefacts are
present.
Table 4-16 VCR Test - Recommended
Step Task Expected Result(s)
1
Connect a Mitsubishi 9600 VCR and video converter to
the system, restart the system and load an SVHS
cassette.
Verify that you follow all the procedure instructions
described in Connecting the VCR on page 3-48.
2
Connect an external monitor to the VCR S-Video or
Video output.
3
Connect a 3S-RS probe. Reset the VCR counter and
press the VCR Record button.
Verify that the recording indicator (red dot) is displayed
on the VCR and the tape counter is running on the
VCR.
4
Activate CW Mode and record 2 minutes of heart
scanning or a similar effect using a phantom.
Record 2 minutes of CW scanning.
5
Stop recording and rewind the cassette. Press the Play
button on the VCR.
The recorded Image is played back.
6
Verify that in CW Mode both the left and right audio
channels are working, and that the played-back audio
is similar to the recorded audio in both the left and right
speakers, when applicable.