GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-29
4-3-2-2 B/W and Color Printer Test
4-3-2-3 Color Printer Test
Table 4-17 B/W Printer Test - Recommended
Step Task Expected Result(s)
1
Scan a phantom or suitable interface in 2D-Mode.
2
Press the Freeze button.
The image scanning stops with the latest picture frozen
on the screen.
3
Press the Print button.
Evaluate the print quality. Adjust the brightness and/or
contrast of the B/W printer, if required. The image
displayed on the screen is printed on the B/W or color
printer, depending upon the key assignment
configuration.
Table 4-18 Color Printer Test - Recommended
Step Task Expected Result(s)
1
Scan a blood vessel in 2D-Mode or a phantom
simulating a similar CFM effect, then activate CFM to
see color.
2
Press the Freeze button.
3
Press the Print P1 button.
Evaluate the color print quality. Adjust the brightness
and/or contrast of the Color printer, if required.