GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-3
Section 8-2
External Component Replacement Procedures
8-2-1 Bearing Handle Replacement Procedure
NOTE: The term "Bearing Handle" relates to the Vivid i n/ Vivid q N portable carrying handle.
8-2-1-1 Tools
Use the appropriate flat and Phillips-type screw drivers as indicated in the Bearing Handle replacement
procedure.
8-2-1-2 Time Required
10 minutes
8-2-1-3 Preparations
None
8-2-1-4 Bearing Handle Removal Procedure
1) Close the Vivid i n/ Vivid q N LCD cover, making sure both front latches are securely fastened
(Figure 8-1).
2) Release the captive screw on either side of the Bearing Handle (there are two screws that secure
Figure 8-1 Closing Vivid i n/ Vivid q N LCD Cover
LatchLatch