GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-32 Section 8-3 - Internal Component Replacement Procedures
8-3-2 TR32 and RFI Boards Replacement Procedure
NOTE: When replacing the TR32 Boards, both boards must be of the same version. Always verify this by
examining the version number stated on the board, as shown in Figure 8-36, below.
CAUTION: When removing the boards, ESD may cause damage to a board. Always have the ESD
hand strip connected to the machine chassis and to your hand.
8-3-2-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-3-2-2 Time Required
15 minutes
8-3-2-3 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
Figure 8-36 TR32 Board Verification (Version 12)
TR32 Board
Version (V12)
Verification