DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-9
8-2-4 Control Panel and Keyboard Replacement Procedure
8-2-4-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the following Control Panel and
Keyboard (and sub-component) replacement procedures.
8-2-4-2 Time Required
10 minutes
8-2-4-3 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-2-4-4 Control Panel and Keyboard Removal Procedure
1) Remove the bearing handle, as described in the Bearing Handle Removal Procedure on page 8 - 3.
2) Place the Vivid i n/Vivid q N portable ultrasound scanner face down and remove the battery as
described in the Battery Removal Procedure on page 8 - 5.
3) Loosen and remove each of the 9 screws that fasten the Control Panel and Keyboard onto the
Bottom Assembly, as indicated in Figure 8-13 below.
4.) Turn the Vivid i n/ Vivid q N over, release the front latches and open the LCD Display cover.
5) Carefully lift the Control Panel and Keyboard upwards from the front, then release it from the rear;
Figure 8-13 Removing the Fastening Screws from the Bottom Assembly
not captive
(do not
This screw
lose!)