GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-10 Section 8-2 - External Component Replacement Procedures
remove it from the Vivid i n/ Vivid q N and place it next to the system, as shown in Figure 8-14.
NOTE: When removing the Control Panel and Keyboard during other replacement procedures, Step 6 will only
be required if indicated.
6) Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and
disconnect the control panel/keyboard cable from the system as shown in Figure 8-15.
Figure 8-14 Removing the Control Panel and Keyboard
Figure 8-15 Disconnecting the Control Panel/Keyboard Cable