GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-53
8-3-8 LCD Display Replacement Procedure
8-3-8-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Display replacement
procedure.
8-3-8-2 Time Required
15 minutes
8-3-8-3 Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-8-4 LCD Display Removal Procedure
1) Remove the bearing handle, as described in the Bearing Handle Removal Procedure on page 8 - 3.
2) Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal
Procedure on page 8 - 9.
3.) Remove the LCD Display Frame as described in the LCD Display Frame Removal Procedure on
page 8 - 27.
4.) Loosen and remove the four screws that fasten the LCD Display to the Rear Cover & Latch
Assembly, as shown in shown in Figure 8-62 below.
5.) Disconnect the LCD Flex ribbon cable from the BEP, as shown in Figure 8-63 below.
Use a flat screwdriver to carefully release the cable from the connector.
Figure 8-62 Removing the Screws from the LCD Display
Figure 8-63 Disconnecting the LCD Flex Ribbon Cable