EasyManua.ls Logo

GE Vivid i n - Image Quality Page; Calibration Page

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-151
7-7-4 Image Quality Page
When the Image Quality page is selected, as shown below, you can verify and calibrate image quality.
Figure 7-118 Image Quality Page
7-7-5 Calibration Page
Figure 7-119 Calibration Page
Note: This page is not populated in this version.
When the Calibration page is selected (as shown in Figure 7-119), although not populated, you can
perform different calibration procedures, as described in Chapter 6 - .
For additional calibration procedures, refer to the following sections:
DC Offset Calibration - see DC Offset Calibration on page 7 - 25.
DC Offset Verification - see DC Offset Verification on page 7 - 28.

Table of Contents

Related product manuals