GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-21
8-2-6 LCD Display Cover Hinges Replacement Procedure
NOTE: This section describes the replacement procedure for the LCD Display Cover hinges only
(Part No. S2381805-2; one pair of hinges).
8-2-6-1 Tools
Phillips screwdriver; Loctite 242 glue
8-2-6-2 Time Required
30 minutes
8-2-6-3 Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-2-6-4 LCD Display Cover Hinges Removal Procedure
1.) Remove the Control Panel and Keyboard, as described in the Control Panel and Keyboard
Removal Procedure on page 8 - 9.
2.) Disconnect the LCD Flex ribbon cable from the BEP, as shown in Figure 8-24.
Use a flat screwdriver to carefully release the cable from the connector.
3.) Loosen and remove the 2 screws from each of the LCD hinges, as shown in Figure 8-25.
Figure 8-24 LCD Flex Cable Connected to BEP
Disconnect
LCD Flex Cable
Here