GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-20 Section 8-2 - External Component Replacement Procedures
5) Return the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation
Procedure on page 8 - 11.
6) Return the Bearing Handle as described in the Bearing Handle Installation Procedure on page 8 - 4.
7) Reconnect the AC adapter and all cables to the Vivid i n/Vivid q N portable ultrasound scanner.
8.) Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23 - only if the Vivid i n/ Vivid q N system
is currently mounted on the SafeLock cart
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
- Full System Test on page 7 - 10