DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-80 Section 8-4 - SafeLock Cart Components Replacement
8-4-11 USB Board Replacement Procedure
8-4-11-1 Tools
Appropriate socket wrench
8-4-11-2 Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-11-3 USB Board Removal Procedure
1) Turn OFF power to the SafeLock Cart and disconnect the AC power cable from the AC input
assembly at the base of the cart (rear).
2) Remove the Upper Cover (Rear), as described in the Upper Cover (Rear) Removal Procedure on
page 8 - 64.
3) Disconnect the P6 cable and USB connector from the USB board, as shown in Figure 8-93.
4) Remove the upper and lower bolts that attach the USB board to the upper frame, as shown in
Figure 8-93.
5) Remove the USB board from the SafeLock Cart.
8-4-11-4 USB Board Installation Procedure
1) Return the replacement USB board to its position on the SafeLock Cart chassis and secure it with
the upper and lower bolts previously removed - see Figure 8-94.
2) Reconnect the P6 cable to the P6 connector on the USB board.
3) Reconnect the USB cable to the USB board.
4) Return the Upper Cover (Rear), as described in the Upper Cover (Rear) Installation Procedure on
page 8 - 64.
Figure 8-93 Removing the USB Box Cover