GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-123
8-5-5 Formatting and Partitioning the Hard Disk - Manual Procedure
NOTE: The procedure described below will create a dedicated partition for storing the ghost image of the
operating system and application software.
If partition C:\ has been stored by using the Rollback tool described in the Software Roll-back
Procedure (see page 8-139), it will be DELETED after performing the following procedure.
1) Insert the Vivid i n/ Vivid q N software installation CD in the DVD\CD-R and close the tray.
2) Hold down the ON/OFF button on the control console for 3 seconds.
The system powers ON and when system boot-up from the CD is complete, the following options
are displayed in the DOS screen, prompting you to select the required option:
3) Select option: 1: Vivid_i Full Clean Install (System + Application Software)
A menu will appear prompting for selection of the required option:
4) Select option B Service Tools.
5) When prompted for the password, type geulsservice and press <Enter>.