EasyManua.ls Logo

GE Vivid i n - Blue Tint 2 Adjustment Procedure

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 6 - Service Adjustments 6-5
6-2-5 Blue Tint "2" Adjustment Procedure
NOTE: Systems with BT11 installed software have a Blue Tint "2" option that enables blue tint enhancement.
This option can be switched off if preferred by the user.
NOTE: This option applies to the following LCD monitor installations:
System: Vivid i n
Monitor: BOE HYDIS-1
NOTE: On older Vivid i/q systems where the IDTECH-1 monitor is supplied, the Blue Tint "2" option is not
applicable.
NOTE: The Blue Tint "2" option is incompatible with IDTECH-1/2 monitors, therefore, on these systems, the
Blue Tint "2" option must be turned off.
To identify which monitor is installed on the system:
1.) Log on as Administrator.
2.) Press Config > About > and select the HwVersion tab.
The monitor type appears under Module Name and the part number indicates the monitor type
- see Figure 6-6.
Figure 6-6 Identifying Monitor Type

Table of Contents