GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-74 Section 8-4 - SafeLock Cart Components Replacement
8-4-9 Peripheral Power Cable Replacement Procedure
8-4-9-1 Tools
Appropriate Phillips screwdriver and wire cutters.
8-4-9-2 Preparation
1) Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
2) Remove the Upper Cover (rear) and Lower Cover (rear) of the SafeLock Cart as described in Upper
Cover (Rear) Removal Procedure on page 8 - 64 and Lower Cover (Rear) Removal Procedure on
page 8 - 71.
8-4-9-3 Peripheral Power Cable Removal Procedure
1) Disconnect the Peripheral Power cable from the AC Distribution Assembly.
2) Cut any cable ties that secure the cable.
3) Disconnect all grounding connections.
4) Remove the Peripheral Power socket by squeezing the plastic clip holders inwards
(see Figure 8-85).
Gently push the power socket inwards to release it.
5) Remove the power socket with its cable.
Figure 8-85 Peripheral Power Socket
Plastic clip holders (right side)