DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 3 - System Setup 3-11
3-4-2 Unpacking the Wooden Transportation box
After completing a visual inspection of the Transportation box and Drop/Tilt indicators, proceed
unpacking as follows:
1.) Remove sealers and unlock two locking levers at two sides of the Box as shown in Figure 3-6 on
page 3-9
2.) Remove the packing slip (shipping consignment note) detailing the contents of the Transportation
Box, keep close to hand and be ready to mark the check list.
3.) Remove the surface packing material (paper pad, silica gel, etc.). Continue to remove additional
packing material as applicable during the following procedure steps.
4.) Carefully remove the box containing the probes.
5.) Take out the package containing the following accessories: cables (AD/DC; ECG; Network),
Isolation Box, and software CD.
6.) Verify the content against the Delivery Note.
7.) Remove each of the boxes (one, two, or more, depending on options ordered) containing the
peripherals and Vivid i n and Vivid q N.
8.) Carefully remove the Vivid i n and Vivid q N ultrasound unit from the box and place at a flat and
stable location- refer to Figure 3-8 on page 3-11.
Figure 3-8 The Vivid i n/ Vivid q N