EasyManua.ls Logo

GE Vivid i n - ECG Module; Overview

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 5 - Components and Function (Theory) 5-39
Section 5-8
ECG Module
5-8-1 Overview
The Vivid i n/ Vivid q N ECG Patient I/O module is designed to enable acquisition of the following
signals: ECG and external ECG. The different pins on the ECG input connector are specifically
assigned to support either External ECG or Internal ECG:
The Internal ECG module supports acquisition of two bipolar ECG channels: leads I, II.
The External ECG module supports acquisition of a single bipolar ECG input channel ranging
from -1 to +1 volt allowing amplification of 1000. The minimal allowed impedance is 10 M ohm.
The ECG module is located under the keyboard assembly (see Figure 5-6 on page 5-9). All power and
communication is supplied to the ECG module via the USB hub located in the keyboard module as
shown in Figure 5-29 below.
Figure 5-29 ECG Patient I/O Module
A general block diagram of the ECG Patient I/O Module is shown in Figure 5-30 below.
Figure 5-30 ECG Patient I/O Module - Block Diagram
Rear
Front
Isolated
Grounded
USB
Cable
ECG Input
3 Leads
Part
Part

Table of Contents

Related product manuals