DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
Section 10-3
Maintenance Task Schedule
10-3-1 How Often Should Care & Maintenance Procedures be Performed?
The Customer Care & Maintenance Task Schedule (provided in Table 10-2 on page 10-5) specifies how
often the Vivid i n/ Vivid q N ultrasound scanner should be serviced, and outlines items requiring special
attention.
NOTE: It is the customer’s responsibility to ensure the Care and Maintenance procedures are performed on the
Vivid i n/ Vivid q N ultrasound scanner as scheduled in order to retain the high levels of safety,
dependability, and system performance.
Your GE Medical Systems Service Representative has an in-depth knowledge of your
Vivid i n/ Vivid q N ultrasound scanning system and can best provide competent, efficient service.
Please contact us for details of your nearest GE Medical Systems Service Representative, and/or a
quotation for services by our qualified professional FSEs.
The service procedures and recommended intervals shown in the Customer Care & Maintenance Task
Schedule assumes that you use your Vivid i n/ Vivid q N scanner for an average patient load (10-12
patients per day) and that you do not use it as a primary mobile unit which is transported between
diagnostic facilities.
NOTE: If conditions exist which exceed typical usage and patient load, it is strongly recommended to increase
the maintenance frequencies.