EasyManua.ls Logo

GE Vivid i n - Page 707

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 10 - Care and Maintenance 10-5
Table 10-2 Customer Care & Maintenance Task Schedule
Service at Indicated Time Daily Weekly Monthly Annually Notes
Clean Probes •* * or before each use
Clean Probe Holders
Inspect AC Mains Cable *• *Mobile Unit Check Weekly
Inspect Cables and Connectors
Clean Console
Clean Monitor
Console Current Leakage Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Peripheral
Current Leakage Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Surface Probe
Current Leakage Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Endocavity Probe Current Leakage
Checks
As prescribed in the Probe Manual.
Transesphongeal (TEE) Probe
Current Leakage Checks
As prescribed in Probe Manual.
Surgical Probe
Current Leakage Checks
As prescribed in Probe Manual.
Measurement Accuracy Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Probe Checks
Also after corrective maintenance,
or as required by your facility’s QA program.
Functional Checks
Also after corrective maintenance,
or as required by your facility’s QA program.

Table of Contents

Related product manuals